ASL Analysis Guide

BASIL perfusion image

Perfusion Quantification using Arterial Spin Labelling MRI

Arterial Spin Labeling (ASL) MRI is a non-invasive method for the quantification of perfusion. This guide introduces the essential concepts you need to understand when acquiring and analysing ASL data, as well as more advanced analysis techniques you might exploit when aiming for higher accuracy and senstivity when using ASL in a study.

This guide complements the information contained in the Introduction to Perfusion quantification using Arterial Spin Labelling (OUP, 2017). You can perform the analyses shown in the guide using the BASIL toolbox by following the instructions in the Tutorials section of this website. Alternatively, similar practical activities are also available on neuroimagingprimers.org.

The guide includes five pre-recorded video lectures that are used in the BASIL course (an FSL mini-course).

Preparatory materials

If you are new to Neuroimaging or to FSL tools you might find the following resources helpful before looking at the material on ASL.

If you are not particularly familar with MRI you might like to read a Short Introduction to MRI Physics for Neuroimaging available via the neuroimagingprimers.org website.

The FSL course provides a wide range of videos and practical exercises. Whilst you do not need to have looked at any of this before reading this guide, you might find some of it helpful background for topics that will be covered (especially for the group analysis section).

We would recommend familarity with:

  • Section 1: Image Registration and Distorition Correction (videos 1-5 are all helpful)
  • Section 2: Segmentation and Structural Analysis (video 7 on FAST)
  • Section 4: Statistics & Task fMRI: Inference and Group Analysis (videos 15-20 provide useful background on group analysis)

Introductory video

In this video we examine what perfusion is and why we might be interested in measuring it in the brain. We briefly consider alternative established perfusion imaging techniques to understand when and where we might use ASL. Finally, we introduce the BASIL toolbox, avaiable as part of FSL.

ASL acquisition

There are various parameters associated with an ASL acquisition. Some of which are important for the quantification/analysis process. Whilst you do not necessarily need to know all of the details of an ASL acquisition to be able to quantify perfusion image, in this section we provide an overview of the key parmeters when designing an ASL protocol and that you need to know for sucessful analysis.

pcASL or pASL

The labelling may either have been pseudo-continuous ASL (pcASL, most common) or pulsed ASL (pASL). There are important differences between these two forms of ASL that affect the quantification, since they determine how the blood-water is labelled.

Post-label delay(s)

After labeling, a delay is left for the labeled blood-water to travel into the brain. For pcASL this is called the Post-Label Delay (PLD) and is the time from the end of the label duration (see Label duration) until imaging. For pASL the labeling process is instantaneous and it is more common to refer to the inversion time (TI). Note that we can define a common delay measure: the inflow time (also TI), the time from the start of labeling. This is identical to the inversion time for pASL, but for pcASL is the sum of the label duration and PLD:

pcASL: TI = PLD + Label duration

It is quite common to meet ASL data with multiple repeats/measurements (and thus volumes in the resulting images) that all have the same PLD (or TI) - single delay ASL. It is, however, possible to use a range of different PLD in an acquisition in an attempt to extract more information, or achieve a better SNR - multi-delay (multi-PLD) ASL.

Label duration

The label (or bolus) duration is an important measure of how much labeled-blood water has been delivered to the tissue and is thus important for quantification. For pcASL the value is set by the sequence and thus is something you need to define/know. It is quite common to use a 1.8 second (or longer) label duration with pcASL.

In principle in pASL the label duration is unknown (a spatial region is labeled instead of a known duration of flowing blood). You may find that your pASL acquisition is using Q2TIPS or QUIPSSII, in which case the label duration has been set using extra pulses. Quite often the value of label duration can then be determined from the associated parameter, often called TI2 - a value of 0.7 or 0.8 seconds would be quite normal. Where the label duration is genuinely unknown (e.g. a FAIR pASL acquisition), it is possible to estimate it as long as the data is multi-TI. It is possible even with Q2TIPS/QUIPSSII that the duration will be shorter than expected due to high flow in the labelling region.

Readout

A variety of readouts can be combined with ASL labelling to acquire ASL data. The important distinction is between (multi-slice) 2D and 3D readouts, since in the former the later time of acquisition of mroe superior slices needs to be acocunted for in the quantification.

Background Suppression

It is common for background suppression to be applied in an ASL acquistion to suppress signal not associated with labeled-blood water (static tissue signal) and reduce artefacts arising from motion.

Analysis of ASL data

The very simplest analysis of ASL data requires the subtraction of label and control images in the data to produce a perfusion weighted image. With the addition of kinetic model inversion and calibration (requring calibration data acquired as part of the ASL dataset) it is possible to produce quantitative perfusion images with conventional usings of ml/100g/min.

Subtraction

Central to ASL analysis is the subtraction of label and control images. Both label and control images will contain some signal from brain tissue - called the static tissue signal (this is true even if background suppression has been used to reduce this contribution). Subtraction of the label-control pair reveals the contribution from labelled blood-water. This image is often referred to as the difference image and is perfusion-weighted, which means it reflects the perfusion in each voxel, but the intensity value in each voxel does not alone provide an absolute measure of perfusion.

To go beyond the perfusion weighted image, and generate quantitative voxelwise measures of perfusion with values in the typical units of ml/100 g/min, we need to use the kinetics of ASL.

Kinetic Model Inversion (Kinetic Modelling)

The voxel intensity in an ASL difference image is directly related to the labelled blood-water. More accurately, it relates to the amount of labelled blood-water that has accumulated in the voxel in the time between creation of the label and the collection of a brain image. This means that it is a measure of delivery and thus perfusion (rather than blood volume or blood flow rate). To be able to say how much labelled blood has been delivered, and thus what the perfusion is, it is necessary to describe the delivery process, as well as what happens to the labelled blood once it has been delivered. This is achieved by means of a kinetic model.

At its very simplest the kinetic model for labelled blood-water in an ASL study needs to account for the delivery of a finite duration (the label duration) of labelled blood-water into the voxel where it accumulates. At the same time as it is being delivered, the label is also decaying away. THe tracer decays at a rate defined by the T1 time constant, which is of the order of a second in the brain at typical MRI field strengths. The kinetic model allows the relationship between the signal and perfusion to be expressed as an equation and this can be rearranged to give an equation that takes signal magnitude and returns perfusion, or fit to the data using optimisation techniques.

Calibration

The ASL calculation relies on knowledge of the tracer concentration, strictly the quantity called the equilibrium magnetization of arterial blood, which will vary between individuals and other MRI-related factors (e.g. the main magnetic field strength). The simplest approach for estimating this parameter is by the acquisition of a separate proton-density-weighted image. This can be converted to a measure of arterial magnetization by accounting for the relative density of hydrogen nuclei in tissue and blood (the partition coefficient). Various corrections can be performed where the calibration image is not a pure proton-density weighted image, e.g., where it has a (realtively) short repetition time.

Further Quantification of ASL data

For single delay ASL data kinetic model inversion is relatively trivial and solutions to the standard model have been described in the literature. However, there are various advantages to aquiring ASL data at multiple times post-inversion and fitting the resultant data to a kinetic model. This permits problems in perfusion estimation associated with variable arterial transit time (ATT) to be avoided, since this becomes a parameter of the model whose value is determined from the data. ATT can also be a valaube parameter (describing the passage of blood throught the vasculature) in its own right.

The model fitting can be performed by a variety of (non-linear) regression techniques, inlcude two step processes that or least squares algorithms. BASIL uses a (fast) Bayesian inference method for the model inversion, this provides a number of advantages:

  • Voxel-wise estimation of perfusion and ATT along with parameter variance (allowing confidence intervals to be calculated).
  • Incorporation of natural varaibility of other model parameters, e.g. values of T1, T1b and labeling/bolus duration.
  • Spatial regularization of the estimated perfusion image.
  • Correction for partial volume effects (where the appropriate segmentation information is available).

Spatial regularization

BASIL can apply a spatial regularisation to the estimated perfusion image and this is highly recommended. This exploits the fact that neighboring voxels are likely to have similar perfusion values, i.e. perfusion variation in the brain is relatively smooth. It brings the advantages associated with the more common pre-processing step of spatially smoothing the data. However, unlike smoothing the data it correctly preserves the non-linear kinetics exploited by the perfusion estimation. It is also adaptive, so that in regions where the data does not support the use of smoothing the perfusion image will not be smoothed.

Group analysis using ASL data

In a study ASL data acquired in individuals can be combined to examine differences or changes in perfusion (or ATT). Group analyses using ASL are similar to that used for other neuroimaging modalities, e.g. BOLD fMRI. I this section we consider specific issues that relate to ASL data, including acheiving good alignment between subjects (registration), the influence of the partial volume effect on computing mean grey matter perfusion values, and what we can do with quantitative measures.

Registration

Registration of ASL data to the structural image is difficult since the images are low resolution and with limited contrast. The most robust approach appears to be to use the perfusion (or perfusion weighted image) since this has greater tissue contrast and is a closer match to a T1-weigthed image than raw ASL (control/label) images. You should ALWAYS inspect the results of registration to determine whether it has been effective.

By default in BASIL registration is carried out in multiple steps using the perfusion image directly after the kinetic model inversion, an intial registration having already been done using the raw (undifferenced) ASL data. BASIL now exploits the BBR cost function for registration since this exploits the boundary between grey and white matter seen in the perfusion images. It is possible use alternative registration strategies and BASIL always produces images in the native space of the data, so that registration can be revisted at a later point.

Advanced Analysis

In the previous sections we have consider what is needed to get a quantative perfusion image out of ASL data. There are a series of additional techniques that can be used to improve the quality and potentially the interpretability of the results. Whilst these techniques are ‘advanced’, since they go beyond the minimum steps outlined in earlier sections, they are by no means necessarily complicated to perform in practice (being built into BASIL).

Correction for Motion, Distortion and Subtraction Artefacts

Strategies used in other neuroimaging modalities to correction for motion and distortion can also be used with ASL data. A particular source of artefacts for ASL arising in the subtraction of label and control images, giving rise to spurious non-perfuion signal components due to motion related differences. Various (and a growing number) of strategies exist to compendate for these.

Arterial (macrovascular) contribution

There can arise signal from labeled arterial blood in the region of major vessels in ASL data. This is most common in data with short PLD (<1.5 s) or in subjects with particular prolonged ATT.

In single PLD ASL data you will need to examine the perfusion images for signs of arterial contaimination (see the ‘White Paper’ for an example of this). This can also be an issue in patients with vascular diseases, where slow flow and thus long ATT are expected and thus longer PLD might be beneficial

For multi delay data the arterial signal can be accounted for by modelling this arterial componen, something included in BASIL by default. When the arterial component is included in the analysis then a further parameter, the arterial blood volume, is available in the output images.

Partial volume correction

The low resolution of ASL data typically means that there is substantial partial voluming of grey (GM) and white matter (WM), plus CSF too. Since GM and WM have very different kinetics (WM tends to have lower perfusion and longer arterial transit time) a normal analysis will provide a perfusion value that is a weighted combination of the two tissue types. Partial Volume Correction attempts to automatically correct for the different tissue type using separately supplied esatimtes of the partial volumes of the tissues. BASIL can do this automatically as long as you supply a structural image that has been already been processed using fsl_anat (or if you supply suitable partial volume estimate images).

T1 values

T1 values are important to the kinetic model inversion and should be chosen based on the field strength that data was acquired at, consideration might also need to be taken of the subject in which analysis is being carried out. BASIL by deafult takes values for 3T and assumes for the tissue only a grey matter value, unless partial volume correction is applied when separate grey and white matter values are specified. By deafult a separate value for the T1 of blood is used unless operating in ‘white paper’ mode, where the blood T1 value is also used for the tissue.

Commonly it is assumed that T1 values are fixed across the brain in the quantification. However, these value are not absolutely certain and may well vary across the brain and between individuals. BASIL can take this into account by inferring on T1 values, you should still, however, set sensible expected values.

ASL variants

You are most likely to be pcASL data in practice. There are variuos other variants of ASL which bring particular advantages, a summary of some notable variants is provided here for reference.

Hadamard/Time-encoded ASL

This is a form of pcASL where the labelling performed via a series of sub-labels with shorter duration. Individual volumes in the ASL acquisition will vary whether for given periods during the label duration labeling is actually taking palce or not. This is normally done accoridng to a specific scheme that means that after decoding it is posisble to recover multi-PLD data that appears as if it has been collected with a PLD equal to the sub-label duration. Even more advanced versions vary the sub-label durations.

To analyse this data you can first perform the decoding step to reveal the multi-PLD data. Thereafter this can be used in BASIL (and associated tools) treating the data as label-control subtracted and specifying the relevant (sub-) label duration and PLDs.

QUASAR

This is a special version of pASL which combines data with and without vascular signal suppression. QUASAR can be used to separate signal from tissue and macrovasular contamination. It is possible using QUASAR to isolate the macrovascular signal and thus estimate an arterial input function, which enables ‘model-free’ deconvolution. QUASAR uses a Look-Locker readout to achieve sampling of different TIs.

Analysis using both ‘model-based’ and ‘model-free’ methods are provided in the QUASIL tool, a version of BASIL optimised for QUASAR data.

Turbo-QUASAR

This is a form of pASL where multiple sub-boluses are created using a series of labelling pulses. It is a variant on QUASAR ASL. The total effective bolus duration is the summation of the duration each sub-bolus, which is equal to the time between each inversion time (TI) of the Look-Locker readout under normal circumstances where the flow velocity of the arterial blood is about 25cm/s. In conditions where the flow velocity is significantly different from this value, an estimation of the flow velocity is needed from a separate phase contrast MR data. Subsequently, the effective bolus duration can be estimated from the flow velocity information.

To analyse Turbo-QUASAR in BASIL, you can the TOAST command line tool.

Further Reading

To learn more about ASL, acquisition choices, the principles of analysis and how perfusion images can be used in group studies you might like to read:

Introduction to Perfusion Quantification using Arterial Spin Labelling, Oxford Neuroimaging Primers, Chappell, MacIntosh & Okell, Oxford University Press, 2017.

Online examples are availble to go with this primer using the BASIL tools. These can be found on the Oxford Neuroimaging Primers website: http://www.neuroimagingprimers.org

The following book reamins a good introduction to functional imaging including perfusion using ASL:

Introduction to Functional Magnetic Resonance Imaging: principles and Techniques. Buxton, Cambridge University Press, 2009.